Instrument No.20103104-9004Specification
ManufactureApplied Photophysics LtdModelChirascan V100
Made inUnited StatesClassification No010411
Location4#220Manufacture date11/09/2019
Purchase Date15/7/2019Internet access time1/11/2019
Main Technical specifications
1. Light source: 150W xenon lamp (UV-visible), halogen lamp (NIR), mercury xenon lamp (applicable to stopped-flow unit),
2. Detector: 170nm-1200nm, APD detector; 850-1700nm, InGaAs detector.
3. Wavelength accuracy: ±0.1nm (160 ~ 500nm) ±0.5nm (500 ~ 1700nm)
4. Adjustable range of spectral bandwidth: 0 ~ >25nm.
5. Stray light: less than 3ppm (200nm).
6. CD noise level:
0.050 m @ 175 nm
0.020 m @ 180 nm
0.030 m @ 200 nm
0.03 mdeg @ 750 nm
0.05 mdeg @ 1000 nm
(detection conditions: 2 second oversampling, 1 nm bandwidth)
7. Measurement range of CD: not less than ±9000 m°
8. Scanning speed: 1-10,000 nm/min
9. Electronic temperature control system (including cyclic bath and magnetic stirring) : -40℃-150℃, accuracy: ± 0.1℃
10. Spectral scanning together with temperature continuous and stepped ramping to study thermodynamic process (Tm, Δ H)
11. Fluorescence polarization/anisotropy detection module: the wavelength range not narrower than 185-850nm.
12. Fast kinetics stopped flow device
Mixing mode: contains two or three samples mixing;
Quench stopped flow device: samples can be recovered for further analysis;
Minimum dead time: better than 1.1 ms (20-micrometer sample, water-water system as the standard);
13. Solid sample transmission measurement accessory
14. Solid sample reflection integrating sphere accessory.
15. Array fluorescent CCD detection.
Main functions and features
Functions:
1. Study the secondary and tertiary structures of biological macromolecules;
2. Study the thermal stability of biological macromolecules and organic polymers;
3. Study on chiral stereoscopic structures of various chiral compounds;
4. Study the millisecond rapid-reaction kinetics mechanism of intermolecular interaction (reaction speed, intermediate transition state existing for a short time), etc.
Features:
Effective detectable spectral range: 175-1700nm;
Sample types: liquid, solid (tablet, film, coating, powder), gas samples, please inquire for details.
Working mode: Static, Kinetics.
A variety of light sources are available.
Services available:
1. Provide static circular dichroic CD detection;
2. Provide reaction kinetics analysis;
3. Provide the use of the instrument;